Further Assessment of Forward Pressure Level for in Situ Calibration
Document Type
Article
Publication Date
12-1-2011
Abstract
Quantifying ear-canal sound level in forward pressure has been suggested as a more accurate and practical alternative to sound pressure level (SPL) calibrations used in clinical settings. The mathematical isolation of forward (and reverse) pressure requires defining the Thvenin-equivalent impedance and pressure of the sound source and characteristic impedance of the load; however, the extent to which inaccuracies in characterizing the source and/or load impact forward pressure level (FPL) calibrations has not been specifically evaluated. This study examined how commercially available probe tips and estimates of characteristic impedance impact the calculation of forward and reverse pressure in a number of test cavities with dimensions chosen to reflect human ear-canal dimensions. Results demonstrate that FPL calibration, which has already been shown to be more accurate than in situ SPL calibration, can be improved particularly around standing-wave null frequencies by refining estimates of characteristic impedance. Better estimates allow FPL to be accurately calculated at least through 10 kHz using a variety of probe tips in test cavities of different sizes, suggesting that FPL calibration can be performed in ear canals of all sizes. Additionally, FPL calibration appears a reasonable option when quantifying the levels of extended high-frequency (10-18 kHz) stimuli.
DOI
10.1121/1.3655878
Montclair State University Digital Commons Citation
Scheperle, Rachel; Goodman, Shawn S.; and Neely, Stephen T., "Further Assessment of Forward Pressure Level for in Situ Calibration" (2011). Department of Communication Sciences and Disorders Faculty Scholarship and Creative Works. 57.
https://digitalcommons.montclair.edu/communcsci-disorders-facpubs/57